|   Clinical Results |
The preliminary clinical results (March 2007) for the custom femoral stem are:
A total of 1200 operated hips
Up to 10 years observation time
Complications:
Three intraoperative fissures (successfully treated with wiring)
Eight patients with moderate thigh pain less than 12 months post operative.
One patient with moderate thigh pain more than 24 months post operative.
Three luxations (fall injuries, stable after reposition)
No loosening
Four revisions due to deep infection (two) and perioprosthetic femoral fracture
University Hospital in Trondheim
The preliminary results (10 years follow up) from non-randomized and randomized studies at Trondheim University Hospital show that Unique® custom stem performs as good as or better than conventional uncemented stem. These studies includes:
function and pain scores
x ray analysis
clinical stability investigations with Radio Stereometric Analysis (RSA)
remodelling investigations with Dual X-ray Absorptiometry (DEXA)
Ullevål University Hospital in Oslo
Preliminary results from randomized studies at Ullevål University Hospital show that Unique custom stem performs as good as or better than a conventional cemented stem. These studies includes :
function and pain scores
x ray analysis
clinical stability investigations with Radio Stereometric Analysis (RSA)